MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/euu365

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2015

Radiofrequency ablation on veno-arterial extracorporeal life support in treatment of very sick infants with incessant tachymyopathy.

Shebani Suhair O, Ng G Andre, Stafford Peter, Duke Christopher

Клинический случайСтатьяПеревод в очереди

Аннотация

AIMS: To evaluate the use of extracorporeal membrane oxygenation (ECMO) in supporting infants who require radiofrequency ablation (RFA) for incessant tachyarrhythmias, with particular emphasis on modifications required to standard ablation techniques.

METHODS AND

RESULTS

Three cases of RFA carried out in infancy on ECMO support were reviewed retrospectively. Two infants with permanent junctional reciprocating tachycardia (PJRT) and one with ventricular tachycardia (VT) presented in a low cardiac output state, owing to cardiomyopathy caused by incessant tachycardia. In each case antiarrhythmic drug therapy caused haemodynamic collapse, requiring emergency ECMO support. Drug therapy on ECMO was not successful. In one patient, the tachycardia was controlled on ECMO with antiarrhythmic drugs, but recurred following ECMO decannulation. Each patient had a successful RFA on ECMO support. Power delivery was low during ablation lesions. In the PJRT cases power as low as 3-5 Watts was effective. In the VT ablation, an irrigated tip RFA catheter was required when cooling remained poor even after temporarily stopping ECMO flow.

CONCLUSION

Extracorporeal membrane oxygenation provides a haemodynamically stable and safe platform for antiarrhythmic drug therapy and RFA in infants with incessant tachyarrhythmias. Once ECMO has been commenced, if the tachyarrhythmia remains difficult to control with antiarrhythmic drugs, RFA should be strongly considered, to avoid the risk of tachycardia recurrence following ECMO decannulation. Power delivery during ablation lesions may be low because of inadequate cooling of the catheter tip. Reducing or stopping flow in the ECMO circuit may not provide adequate cooling and an irrigated tip catheter may be required.

Для цитирования: Shebani Suhair O, Ng G Andre, Stafford Peter, Duke Christopher. Radiofrequency ablation on veno-arterial extracorporeal life support in treatment of very sick infants with incessant tachymyopathy.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2015. DOI: 10.1093/europace/euu365.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.