OMERACT GCA phantom project: validation of a 3D-printed ultrasound training phantom for diagnosis of giant cell arteritis.

Schremmer Tobias, Dejaco Christian, Recker Florian, Aschwanden Markus, Boumans Dennis, Bruyn George A, Chrysidis Stavros, Daikeler Thomas, Nielsen Berit, De Miguel Eugenio, Diamantopoulos Andreas P, Dohn Uffe Moller, Duftner Christina, Germano Giuseppe, Haagsma Cees, Hartung Wolfgang, Hocevar Alojzija, Iagnocco Annamaria, Juche Aaron, Karakostas Pantelis, Karalilova Rositsa, Keller Kresten Krarup, Sarker Borsha, Smith Kate, Tischler Verena, Macchioni Pierluigi, Milchert Marcin, Monti Sara, Mukhtyar Chetan, Naredo Esperanza, Schmidt Wolfgang A, Seitz Luca, Terslev Lene, Wakefield Richard J, Schäfer Valentin Sebastian

СтатьяВалидационное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

Ultrasonography is crucial for diagnosing giant cell arteritis (GCA); however, training opportunities are rare. This study tested the reliability of ultrasonography findings and measurements of the intima-media thickness (IMT) among ultrasonography experts by using phantoms of the axillary (AA) and temporal arteries (TA) created with high-resolution 3D printing.

METHODS

Twenty-eight participants from 12 European countries received eight sets of phantoms of the AA and the superficial TA (including common, frontal and parietal branches), which were examined in a blinded fashion according to a predefined protocol and evaluated based on Outcome Measures in Rheumatology (OMERACT) GCA ultrasound definitions. Due to difficulties with the delineation of the intima-media complex, the parietal branch of the phantoms was modified, and a second round was conducted. The IMT was measured, and phantoms were classified as normal or vasculitic.

RESULTS

In both rounds, the phantoms were correctly classified as normal/abnormal in >81% of cases yielding a Fleiss' kappa of 0.80 (95% CI 0.78 to 0.81) in round 1 and 0.74 (95% CI 0.72 to 0.75) in round 2. IMT measurements revealed an intraclass correlation coefficient (ICC 1.1) of 0.98 (95% CI 0.98 to 0.99) in both rounds. Intrarater reliability was good with a median Cohens Kappa of 0.83 and median ICC of 0.78.

CONCLUSION

The study demonstrated high reliability among ultrasound experts in applying the OMERACT ultrasound definitions for GCA and in measuring the IMT using a 3D-printed phantom of the AA and TA. This phantom could assist clinicians in training to assess the large arteries of patients with suspected or established GCA.

Для цитирования: Schremmer Tobias, Dejaco Christian, Recker Florian, Aschwanden Markus, Boumans Dennis, Bruyn George A, Chrysidis Stavros, Daikeler Thomas, Nielsen Berit, De Miguel Eugenio, Diamantopoulos Andreas P, Dohn Uffe Moller, Duftner Christina, Germano Giuseppe, Haagsma Cees, Hartung Wolfgang, Hocevar Alojzija, Iagnocco Annamaria, Juche Aaron, Karakostas Pantelis, Karalilova Rositsa, Keller Kresten Krarup, Sarker Borsha, Smith Kate, Tischler Verena, Macchioni Pierluigi, Milchert Marcin, Monti Sara, Mukhtyar Chetan, Naredo Esperanza, Schmidt Wolfgang A, Seitz Luca, Terslev Lene, Wakefield Richard J, Schäfer Valentin Sebastian. OMERACT GCA phantom project: validation of a 3D-printed ultrasound training phantom for diagnosis of giant cell arteritis.. RMD open. 2025. DOI: 10.1136/rmdopen-2024-005316.