Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2017
Venous access and long-term pacemaker lead failure: comparing contrast-guided axillary vein puncture with subclavian puncture and cephalic cutdown.
Сравнительное исследованиеСтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Chan Ngai-Yin, Kwong Nim-Pong, Cheong Adrian-Piers
Аннотация
AIMS: Existing data on the relationship between venous access and long-term pacemaker lead failure (PLF) are scarce and inconsistent. We aim to study the hypothesis that contrast-guided axillary vein puncture (AP) is better than subclavian puncture (SP) and similar to cephalic vein cutdown (CV) in the incidence of PLF and the success rate of AP is higher than CV.
METHODS AND RESULTS: The case records of 409 patients with 681 implantable pacemaker leads were reviewed. Two hundred and fifty-two, 217, and 212 leads were implanted via AP, CV, and SP, respectively. With a mean follow-up of 73.6 ± 33.1 months, 20 (2.9%) PLF occurred. Three (1.2%), 5 (2.3%), and 12 (5.6%) PLF occurred in the AP, CV, and SP groups, respectively. On multivariate Cox regression analysis, the only independent predictor for PLF was the use of SP instead of AP (AP vs. SP; hazard ratio: 0.261; 95% confidence interval: 0.071-0.954, P = 0.042). The success rate of CV (78.2%) was significantly lower than those of AP (97.6%) and SP (96.8%) (P < 0.001).
CONCLUSION: Compared with SP, the use of AP but not CV independently predicted a lower risk of PLF. The success rates in achieving venous accesses were similar between AP and SP, but significantly lower for CV. Axillary vein puncture may thus be considered the venous access of choice for pacemaker lead implantation.
METHODS AND RESULTS: The case records of 409 patients with 681 implantable pacemaker leads were reviewed. Two hundred and fifty-two, 217, and 212 leads were implanted via AP, CV, and SP, respectively. With a mean follow-up of 73.6 ± 33.1 months, 20 (2.9%) PLF occurred. Three (1.2%), 5 (2.3%), and 12 (5.6%) PLF occurred in the AP, CV, and SP groups, respectively. On multivariate Cox regression analysis, the only independent predictor for PLF was the use of SP instead of AP (AP vs. SP; hazard ratio: 0.261; 95% confidence interval: 0.071-0.954, P = 0.042). The success rate of CV (78.2%) was significantly lower than those of AP (97.6%) and SP (96.8%) (P < 0.001).
CONCLUSION: Compared with SP, the use of AP but not CV independently predicted a lower risk of PLF. The success rates in achieving venous accesses were similar between AP and SP, but significantly lower for CV. Axillary vein puncture may thus be considered the venous access of choice for pacemaker lead implantation.
Предметные рубрики MeSH
AgedAged, 80 and overAxillary VeinCardiac Pacing, ArtificialCatheterization, Central VenousCatheterization, PeripheralChi-Square DistributionContrast MediaEquipment FailureFemaleHong KongHumansKaplan-Meier EstimateMaleMiddle AgedMultivariate AnalysisPacemaker, ArtificialPhlebographyPredictive Value of TestsProportional Hazards ModelsPuncturesRetrospective StudiesRisk FactorsSubclavian VeinTime FactorsTreatment OutcomeVenous Cutdown