MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1017/s1047951122003547

Медицинская библиотека · Cardiology in the young · 2022

Early and mid-term outcomes of transcatheter closure of perimembranous ventricular septal defects using the Lifetech™ Konar-MF Occluder device (MFO).

Odemis Ender, Kizilkaya Mete Han

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Transcatheter closure of perimembranous ventricular septal defects is challenging. Double disk-designed devices have high adverse event rates; therefore, research into new devices persists. One such devise is the LifetechTM Konar-MF Occluder device (MFO), which is increasingly used. The purpose of this study is to present mid-term results of MFO for transcatheter closure of patients with perimembraneous ventricular septal defect.

PATIENTS AND

METHODS

Records of the 52 patients with perimembraneous ventricular septal defect who had undergone transcatheter closure with MFO were reviewed. Demographic, echocardiographic, and procedure details were investigated. Three years follow-up results were recorded.

RESULTS

We closed the perimembraneous ventricular septal defect in 51 of 52 patients (98%). The femoral venous approach was used in 27 patients (53 %) whilst no arteriovenous loop was established in the remaining patients. No significant procedure-related complication occurred. On day 1 echocardiography, the residual ventricular septal defect rate was 31%. Mean±SD (range) follow-up duration was 36 ± 7,9 (18-54) months. In the follow-up, at 6th months, only four patients had hemodynamically insignificant residual defects. No severe dysrhythmia was detected including complete heart block. A right bundle branch block pattern was seen in one patient.

CONCLUSION

This study showed that MFO is a safe and effective device in the transcatheter treatment of perimembraneous ventricular septal defect's with mid-term follow-up.

Для цитирования: Odemis Ender, Kizilkaya Mete Han. Early and mid-term outcomes of transcatheter closure of perimembranous ventricular septal defects using the Lifetech™ Konar-MF Occluder device (MFO).. Cardiology in the young. 2022. DOI: 10.1017/s1047951122003547.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.