m+Medelevator.RU
← Поиск публикаций
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2025

Biogenic Polymers for Congenital Cardiac Surgery: In Vitro Durability Testing.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Hubrich Julian, Grefen Linda, Zorin Alexandra, Richert Emma, Simon Stefan, Jahren Silje, Hagl Christian, Carrel Thierry, Hörer Jürgen, Herz Christopher, Grab Maximilian, Heinisch Paul Philipp

Аннотация

OBJECTIVES: Currently available materials in congenital cardiac surgery (CCS) are far from optimal because they do not facilitate growth, remodelling, or renewal, resulting in poorer long-term outcomes due to material-related limitations. Bacterial cellulose (BC), a biogenic polymer-based material produced by Acetobacter xylinum, has emerged as a promising alternative exhibiting excellent bio- and haemocompatibility. This study aimed to develop BC specifically for application in CCS by modifying culturing conditions to enhance its biomechanical resilience.

METHODS: Bacterial cellulose was produced according to a standard protocol, and its biomechanical properties were evaluated using inflation pressure testing, thickness measurement, and uniaxial tensile testing. To improve these characteristics, 2 modifications, a change in the growth media composition and an extended incubation time, were implemented and subsequently evaluated in 5 further test series. The long-term durability of BC patches was assessed in a fatigue tester for 20 ± 0.5 million cycles, and potential structural damage was investigated using scanning electron microscopy.

RESULTS: Utilizing the 2 modifications, BC patches demonstrated a capability to reach and maintain pressures exceeding 1000 mm Hg with a durability of 100% (n = 24), compared to the standard 20.8% (n = 24) at 500 mm Hg. The maximum tensile strength was enhanced from 0.311 ± 0.057 megapascal (MPa) to 0.986 ± 0.397 MPa (P < .0001), with a thickness of 3.89 ± 0.95 mm (P < .0001). In the long-term durability testing, patches endured durations equalling 6 months without failure while retaining structural integrity.

CONCLUSIONS: The potential of BC for use in CCS was demonstrated by enhancing its biomechanical properties through culturing modifications, warranting further investigation and development of the biomaterial.