DOI 10.1097/scs.0000000000006821
Accuracy of Patient-Specific Meshes as a Reconstruction of Orbital Floor Blow-Out Fractures.
Аннотация
Computer-aided design and manufacturing (CAD-CAM)-based techniques are developing fast in facial reconstruction and osteosynthesis. Patient-specific implant (PSI) production is already sufficiently fast for everyday use and can be utilized even for primary trauma surgery such as orbital floor reconstruction after blowout fracture. Purpose of our study is to retrospectively analyze the 3-dimensional (3D) success of PSI reconstructions of orbital floor fractures in our unit. The authors analyzed retrospectively a 1-year cohort (n = 8) of orbital floor blow-out fractures that have been reconstructed using virtual surgical plan and CAD-CAM PSI. Postoperative computed topographies of patients were compared to their original virtual surgical plans. The 3D outcome and fitting of the PSI was good in all patients. Mean error for 3D position of the PSI was 1.3 to 1.8 mm (range 0.4 to 4.8 mm) and postoperative orbital volume was successfully restored in all of the patients. Use of CAD-CAM PSI for reconstruction of orbital floor blow out fracture is reliable method and thus recommended.