Pulse oximetry measures a lower heart rate at birth compared with electrocardiography.
Аннотация
OBJECTIVE
To examine the effect of time after birth on heart rate (HR) measured by pulse oximetry (PO) (HRPO) and electrocardiography (ECG) (HRECG).
STUDY
DESIGN
HRECG and HRPO (collected at maximum sensitivity) were assessed in 53 term and preterm infants at birth. ECG electrodes and a PO sensor were attached as soon as possible and HRECG and HRPO were compared every 30 seconds from 1-10 minutes after birth. Data were compared using a Wilkinson signed-rank test. Clinical relevance (eg, HR <100 beats per minute [bpm] was tested using a McNemar test).
RESULTS
Seven hundred fifty-five data pairs were analyzed. Median (IQR) gestational age was 37 (31-39) weeks. Mean (SD) starting time of PO and ECG data collection was 99 (33) vs 82 (26) seconds after birth (P = .001). In the first 2 minutes after birth, HRPO was significantly lower compared with HRECG (94 (67-144) vs 150 (91-153) bpm at 60 seconds (P < .05), 81 (60-109) vs 148 (83-170) bpm at 90 seconds (P < .001) and 83 (67-145) vs 158 (119-176) at 120 seconds (P < .001). A HR <100 bpm was more frequently observed with a PO than ECG in the first 2 minutes (64% vs 27% at 60 seconds (P = .05), 56% vs 26% at 90 seconds (P < .05) and 53% vs 21% at 120 seconds (P < .05). HR by ECG was verified by ultrasound for outflow from a subset of infants.
CONCLUSIONS
In infants at birth, HRPO is significantly lower compared with ECG with clinically important differences in the first minutes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.