MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1017/s1047951122003973

Медицинская библиотека · Cardiology in the young · 2022

Risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus.

Mumtaz Zeeshan A, Sagar Pramod, Sivakumar Kothandam, Mohakud Asish R, Rajendran Monica, Pavithran Sreeja

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Device embolisation is a serious adverse event during transcatheter duct closure. This study analyses risk factors for embolisation.

METHODS

Demographic parameters, echocardiographic anatomy, haemodynamics, and procedural characteristics of consecutive duct closures in a tertiary centre over 8 years were analysed. Procedures complicated by embolisation were compared to uncomplicated procedures.

RESULTS

Fifteen embolisations occurred during 376 procedures. All except one embolisation were in infants. The pulmonary artery: aortic pressure ratio was 0.78 ± 0.22. Embolisation was seen significantly more commonly in Type C tubular ducts. Vascular plugs were more significantly associated with embolisations. Logistic regression analysis showed device embolisation was significantly higher in age group of < 6 months compared to 6-12 months (p = 0.02), higher in those with tubular ducts versus conical ducts (p = 0.003), use of vascular plugs compared to conventional duct occluders (p = 0.05), and in duct closure with undersized devices (p = 0.001). There was no in-hospital mortality. Three patients needed surgical retrieval while others were successfully managed in catheterisation laboratory.

CONCLUSIONS

Device embolisation complicates 4% of transcatheter duct closures, with need for surgery in one-fifth of them. Larger ducts with high pulmonary artery pressures in younger and smaller infants are more often associated with device embolisation. Tubular ducts are more prone for embolisation compared to usual conical ducts. Softer vascular plugs are often associated with embolisations. Intentional device undersizing to avoid vascular obstruction in small patients is a frequent risk factor for embolisation. Precise echocardiographic measurements, correct occluder choice, proper technique and additional care in patients with high pulmonary artery pressures are mandatory to minimise embolisations.

Для цитирования: Mumtaz Zeeshan A, Sagar Pramod, Sivakumar Kothandam, Mohakud Asish R, Rajendran Monica, Pavithran Sreeja. Risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus.. Cardiology in the young. 2022. DOI: 10.1017/s1047951122003973.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.