Comparison of cardiac indices during pregnancy between noninvasive automated device and two-dimensional echocardiography: a longitudinal study.

Chaemsaithong Piya, Chung Man Yan, Sun Qian, Kwan Angel H W, Kubi Appiah, Huang Junhong, Chong Ka Chun, Poon Liona C

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To determine the correlation, precision, mean percentage difference and agreement between cardiac indices (stroke volume [SV], cardiac output [CO], SV index [SVI], and cardiac index [CI]) measured by noninvasive cardiac output monitor (NICOM®, Cheetah Medical, Boston, MA, USA) and 2-dimensional transthoracic echocardiography (2D-TTE) across gestations in Chinese pregnant women.

METHODS

This was a prospective longitudinal study performed in women with singleton pregnancy at 11-14+6 (n = 152), 19-24+6 (n = 152), 30-34+6 (n = 141), and 35-37+6 (n = 103). Cardiac indices, including CO and SV, were obtained by NICOM®, which uses thoracic bioreactance, and 2D-TTE. CI and SVI were calculated from CO and SV adjusted for body surface area. The measurements of cardiac indices obtained using NICOM® were assessed relative to that of 2D-TTE by calculating correlation coefficient, bias, precision, mean percentage difference, and 95% limits of agreement, adjusted for repeated measurements.

RESULTS

Comparison of the SV and SVI measurements by the two approaches showed significant moderate correlation in the first trimester (r = 0.2-0.3; p = .01). Overall, the SV and SVI measurements obtained using NICOM® relative to that obtained by 2D-TTE revealed a bias of -12.1 mL and -6.1 mL/m2 (95% confidence interval [CI]: -44.5 to 20.2 and -24.8 to 12.5), respectively. Comparison of the CO and CI measurements by the two approaches showed significant moderate correlation in the first trimester (r∼0.2; p = .01). Overall, the CO and CI measurements obtained using NICOM® relative to that obtained by 2D-TTE revealed a bias of -0.50 L/min and -0.18 L/min/m2 (95% CI: 2.26-3.27 and -1.79 to 1.39), respectively. Mean percentage difference for all cardiac parameters in all three trimesters were more than 30%.

CONCLUSIONS

In Chinese pregnant women, NICOM® has underestimated cardiac indices (SV, CO, SVI, and CI) compared to that measured by 2D-TTE. The mean percentage differences for all cardiac indices are more than 30%, which are higher than the clinically acceptable limit. Future research is needed to determine whether adjustment factors should be applied to the proprietary algorithms used by the NICOM® for the estimation of SV and CO in the Chinese pregnant population.

Для цитирования: Chaemsaithong Piya, Chung Man Yan, Sun Qian, Kwan Angel H W, Kubi Appiah, Huang Junhong, Chong Ka Chun, Poon Liona C. Comparison of cardiac indices during pregnancy between noninvasive automated device and two-dimensional echocardiography: a longitudinal study.. The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians. 2024. DOI: 10.1080/14767058.2020.1865908.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.