Noninvasive thermal evaluation for shunt failure in the emergency room.

Xu Jordan, Poole Cassie, Sahyouni Ronald, Chen Jefferson

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Ventriculoperitoneal shunts (VPSs) have been the mainstay of treating hydrocephalus since the 1950s. However, shunts have a reported complication rate reaching nearly 50%. Devices have been developed that utilize noninvasive thermal transcutaneous diffusion technology. These shunt evaluation devices measure temperature gradients to detect shunt cerebrospinal fluid flow. We assessed the utility using a thermal diffusion technique to work up shunt failure in the emergency room (ER).

METHODS

This was a retrospective case series at a single medical center ER. We included consecutive patients with possible VPS malfunction who were evaluated with a thermal sensor. The time period of data collection included September 2015-April 2018.

RESULTS

Eight patients were reviewed, including four males and four females. The average age of reviewed patients was 35.1 (+/- ; 16.5). In our case series, three patients had adequate shunt flow as assessed by the shunt evaluation device, and the decision was made to discharge the patient and follow-up in clinic. In two patients, the flow was diminished, but due to other reassuring signs, the patients were still discharged with follow-up. Two patients were noted to have poor flow and were admitted for shunt revision.

CONCLUSION

It is important to determine whether a malfunction is present and whether an intervention is necessary for patients who present to the emergency department with possible symptoms of shunt failure. A thermal sensor is a fast and noninvasive tool in the evaluation of shunt flow and helps determine whether it is safe to send a patient home or intervene appropriately.

Для цитирования: Xu Jordan, Poole Cassie, Sahyouni Ronald, Chen Jefferson. Noninvasive thermal evaluation for shunt failure in the emergency room.. Surgical neurology international. 2019. DOI: 10.25259/sni_324_2019.