VIT-MBRT, a GPU accelerated Monte Carlo tool for investigations on preclinical minibeam radiation therapy.

Luongo N, Crimaldi U, Cerbone L A, Jia X, Lai Y, Clemente S, Oliviero C, Pacelli R, Spinelli A, Fiorino C, Pizzardi S, Russo P, Mettivier G

СтатьяПеревод в очереди

Аннотация

PURPOSE: Minibeam Radiation Therapy (MBRT) is a form of spatially fractionated radiation therapy (SFRT) with electrons, protons, ions or kilovoltage photon beams, aiming to improve the therapeutic window compared to broad-beam irradiation. MBRT with X-rays uses an orthovoltage source and an array of narrow beamlets produced by collimator slits with a typical width of 0.2 mm to 0.7 mm and a center-to-center spacing of 1 to 4 mm. We developed an updated version of gCTD, a fast Monte Carlo (MC) code, initially designed for cone-beam CT imaging, to implement a treatment planning system for photon MBRT in mice within a virtual imaging and dosimetry tool (VIT-MBRT).

METHODS

We compared gCTD, running on a single GPU platform (CUDA environment) with TOPAS, a well validated CPU-based simulation code. We simulated irradiation of a water phantom using a multi-slit tungsten collimator with different aperture width and thickness, producing absorbed dose volume data. An example treatment plan was implemented with the gCTD code, using a voxelized mouse phantom derived from a CT scan and evaluating organ dose-volume histograms.

RESULTS

Comparison with TOPAS simulations showed a good agreement in terms of dose values and peak-to-valley dose ratio (maximum absolute discrepancy of 13% at the phantom entrance surface), with a few percent differences in depth. Simulations with gCTD achieved a 300-fold reduction in computational time with respect to corresponding TOPAS simulations.

CONCLUSIONS

We realized and validated a fast GPU-based MC simulation code for minibeam radiotherapy, as the basis of the MC platform VIT-MBRT for kilovoltage MBRT preclinical treatments.

Для цитирования: Luongo N, Crimaldi U, Cerbone L A, Jia X, Lai Y, Clemente S, Oliviero C, Pacelli R, Spinelli A, Fiorino C, Pizzardi S, Russo P, Mettivier G. VIT-MBRT, a GPU accelerated Monte Carlo tool for investigations on preclinical minibeam radiation therapy.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2026. DOI: 10.1016/j.ejmp.2026.105859.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.