Total Retinal Blood Flow in a Nonhuman Primate Optic Nerve Transection Model Using Dual-Beam Bidirectional Doppler FD-OCT and Microsphere Method.
Аннотация
PURPOSE: We validated noninvasive Doppler-optical coherence tomography (OCT) blood flow measurements against the terminal microsphere method in a surgical induced optic nerve transection nonhuman primate model.
METHODS
In 6 nonhuman primates, total retinal blood flow (TRBF) was measured with a custom-built dual-beam bidirectional Doppler Fourier Domain (FD)-OCT. Peripapillary retinal nerve fiber layer thickness (RNFLT) was measured by Spectralis spectral-domain (SD)-OCT. Measurements were performed every 10 to 15 days before and after unilateral optic nerve transection (ONT) until RNFLT was reduced by more than 40% from baseline. Before the animals were killed, TRBF was measured using the microsphere technique.
RESULTS
A significant correlation between all arterial and venous Doppler OCT TRBF measurements was found in ONT and contralateral control eyes (both P < 0.01, n = 6). The Bland-Altman analysis showed a bias of 0.57 in the ONT group and 0.02 in the contralateral control group. Also, excellent agreement was observed between Doppler OCT and microsphere measurements (P < 0.01, r = 0.976, bias = 0.54). After ONT, TRBF and RNFLT decreased by -51% ± 42% and -44% ± 2% (n = 5), respectively. In the contralateral control eyes, TRBF and RNFLT were unchanged.
CONCLUSIONS
Very good accordance was found between TRBF measurements, obtained with dual-beam bidirectional Doppler FD-OCT and the microsphere method. It also was possible to monitor changes over time in TRBF after ONT with Doppler OCT. These findings highlight the accuracy and potential of noninvasive Doppler OCT to provide valuable information for detecting early changes in ocular disease in future.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.