Expected drainage pressures on extracorporeal membrane oxygenation based on in-vitro modeling of pressure-flow variables.
Аннотация
BACKGROUND
Pressure readings used in extracorporeal membrane oxygenation (ECMO) guide a variety of clinical interventions. However, several factors affect the accuracy of these measurements. We aimed to determine a model to mathematically predict expected pressure readings for different pediatric ECMO cannulae.
METHODS
We assembled an in-vitro circuit simulating those used clinically. Flow (mL/min) was adjusted, and the corresponding pressure drop (mm Hg) across the cannula was documented for various Medtronic Bio-medicus® and Life Support® cannulae. We utilized both water and blood product mixtures of differing hematocrit (Hct) levels - analogous to differing viscosities. Experimentally derived mathematical models were then applied to clinical data.
RESULTS
The relationship between flow and pressure drops across different cannulae and at different Hct levels was graphed and analyzed. At the same flow, larger cannula size corresponded to smaller drops in pressure, and increased viscosity corresponded to larger drops in pressure. Using regression models of the experimental data, we estimated five exponential equations that include Hct and flow to predict the expected pressure drop for different pediatric ECMO cannulae. There was high variability in data among patients, with significant differences between clinically observed and expected pressures.
CONCLUSION
Our models derived from in-vitro experimentation are the first steps towards defining the pressure-flow relationship and validating drainage pressure as a surrogate measure of intravascular volume for critically ill patients on ECMO. Further studies are still needed to refine these models and validate their clinical applicability.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.