DOI 10.1038/s41390-019-0304-0

Медицинская библиотека · Pediatric research · 2019

The point-of-care Bilistick method has very short turn-around-time and high accuracy at lower cutoff levels to predict laboratory-measured TSB.

Boo Nem-Yun, Chang Yun-Fah, Leong Ying-Xin, Tok Zhen-Yu, Hooi Lai-Cheng, Chee Seok-Chiong, Latif Zuraidah Abdul

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

This study aimed to determine the accuracy of a point-of-care Bilistick method for measuring total serum bilirubin (TSB) and its turn-around-time (TAT) against hospital laboratory methods.

METHODS

This prospective study was carried out on 561 term-gestation jaundiced neonates in two Malaysian hospitals. Venous blood sample was collected from each neonate for contemporary measurement of TSB by hospital laboratories and Bilistick. TAT was the time interval between specimen collection and TSB result reported by each method.

RESULTS

The mean laboratory-measured TSB was 194.85 (±2.844) µmol/L and Bilistick TSB was 169.37 (±2.706) µmol/L. Pearson's correlation coefficient was: r = 0.901 (p < 0.001). The mean difference of [laboratory TSB- Bilistick TBS] was 26.48 (±29.41) µmol/L. The Bland-Altman plots show that the 95% limits of agreement (-31.1577, 84.11772) contain 94.7% (=531/561) of the difference in TSB readings. Bilistick has a 99% accuracy and 100% sensitivity to predict laboratory TSB levels of ≥80 µmol/L and ≥360 µmol/L at lower Bilistick TSB levels of ≥55 and ≥315 µmol/L, respectively. TAT of Bilistick TSB (2.0 min) was significantly shorter than TAT (105 min) of laboratory TSB (p < 0.001).

CONCLUSIONS

Bilistick has shorter TAT. The accuracy and sensitivity of Bilistick TSB for predicting laboratory TSB is high at lower cutoff levels.

Для цитирования: Boo Nem-Yun, Chang Yun-Fah, Leong Ying-Xin, Tok Zhen-Yu, Hooi Lai-Cheng, Chee Seok-Chiong, Latif Zuraidah Abdul. The point-of-care Bilistick method has very short turn-around-time and high accuracy at lower cutoff levels to predict laboratory-measured TSB.. Pediatric research. 2019. DOI: 10.1038/s41390-019-0304-0.