Dose optimization in newborn abdominal radiography: Assessing the added value of additional filtration on radiation dose and image quality using an anthropomorphic phantom.

Petit Annie-Lyne, Alwan Rabih, Behr Julien, Calame Paul, Lenoir Marion, Ducou le Pointe Hubert, Delabrousse Éric

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Abdominal radiographs remain useful in newborns. Given the high radiation sensitivity of this population, it is necessary to optimize acquisition techniques to minimize radiation exposure.

OBJECTIVE

Evaluate the effects of three additional filtrations on radiation dose and image quality in abdominal X-rays of newborns using an anthropomorphic phantom.

MATERIAL AND METHOD: Abdominal radiographs of an anthropomorphic newborn phantom were performed using acquisition parameters ranging from 55 to 70 kV and from 0.4 to 2.5 mAs, without and with three different additional filtrations: 0.1 mm copper (Cu) + 1 mm aluminum (Al), 0.2 mm copper + 1 mm aluminum, and 2 mm aluminum. For each X-ray the dose area product (DAP) was measured, the signal-to-noise ratio (SNR) was calculated, and image quality (IQ) was evaluated by two blinded radiologists using the absolute visual grading analysis (VGA) method.

RESULTS

Adding an additional filtration resulted in a significant reduction in DAP, with a decrease of 42% using 2 mm Al filtration, 65% with 0.1 mm Cu + 1 mm Al filtration, and 78% with 0.2 mm Cu + 1 mm Al filtration (p < 0.01). The addition of 2 mm aluminum filtration does not significantly decrease the SNR (p = 0.31), CNR (p = 0.52) or the IQ (p = 0.12 and 0.401 for reader 1 and 2, respectively). However, adding copper-containing filtration leads to a significant decrease in, SNR, CNR and IQ.

CONCLUSION

Adding a 2 mm Al additional filtration for abdominal radiographs in newborns can significantly reduce the radiation dose without causing a significant decrease in image quality.

Для цитирования: Petit Annie-Lyne, Alwan Rabih, Behr Julien, Calame Paul, Lenoir Marion, Ducou le Pointe Hubert, Delabrousse Éric. Dose optimization in newborn abdominal radiography: Assessing the added value of additional filtration on radiation dose and image quality using an anthropomorphic phantom.. Research in diagnostic and interventional imaging. 2024. DOI: 10.1016/j.redii.2024.100045.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.