A Novel Whole Spectrum-Based Non-Invasive Screening Device for Neonatal Hyperbilirubinemia.

Halder Animesh, Banerjee Mukut, Singh Soumendra, Adhikari Aniruddha, Sarkar Probir Kumar, Bhattacharya Arnab Madhab, Chakrabarti Prantar, Bhattacharyya Debasis, Mallick Asim Kumar, Pal Samir Kumar

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

Careful screening of bilirubin level in newborns is mandatory as per American Academy of Pediatrics (2004), to reduce incidents of kernicterus and acute bilirubin encephalopathy. Although, invasive capillary collection of blood and subsequent biochemical test is considered a gold standard for jaundice detection in neonates, transcutaneous bilirubin measurement using various non-invasive instruments is also used sporadically across the globe. The major aim of this study was to develop a non-invasive spectrometry-based technique for measurement of neonatal bilirubin level as an alternative of total serum bilirubin (TSB) test without limitations of other available bilirubinometers.

METHODS

The instrument comprises of a light source and a spectroscopic detector. A light beam from source incident on the neonatal nail plate through optical fibers. The retro reflected light is acquired using the detector. An indigenously developed software is used to acquire and analyze the optical signal and to calculate the bilirubin value. The instrument was calibrated and validated in reference to TSB on 1033 subjects.

MAJOR

RESULTS

The result (r = 0.95 and P < 0.001) indicates a strong correlation between the bilirubin value obtained from our method and TSB. Time variant analysis of the subjects undergoing phototherapy provided a good correlation (r = 0.98). The repeatability test result shows the mean coefficient of variation is less than 5.0%.

CONCLUSIONS

The indigenously developed non-invasive technique successfully detects the bilirubin level in newborns under various physiological conditions with high accuracy and precision.

Для цитирования: Halder Animesh, Banerjee Mukut, Singh Soumendra, Adhikari Aniruddha, Sarkar Probir Kumar, Bhattacharya Arnab Madhab, Chakrabarti Prantar, Bhattacharyya Debasis, Mallick Asim Kumar, Pal Samir Kumar. A Novel Whole Spectrum-Based Non-Invasive Screening Device for Neonatal Hyperbilirubinemia.. IEEE journal of biomedical and health informatics. 2019. DOI: 10.1109/jbhi.2019.2892946.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.