A laparoscopic knot-tying device for minimally invasive cardiac surgery.

Jernigan Shaphan R, Chanoit Guillaume, Veeramani Arun, Owen Stephen B, Hilliard Matthew, Cormier Denis, Laffitte Bryan, Buckner Gregory

Оценочное исследованиеСтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

OBJECTIVE

Intracorporeal suturing and knot tying can complicate, prolong or preclude minimally invasive surgical procedures, reducing their advantages over conventional approaches. An automated knot-tying device has been developed to speed suture fixation during minimally invasive cardiac surgery while retaining the desirable characteristics of conventional hand-tied surgeon's knots: holding strength and visual and haptic feedback. A rotating slotted disc (at the instrument's distal end) automates overhand throws, thereby eliminating the need to manually pass one suture end through a loop in the opposing end. The electronic actuation of this disc produces left or right overhand knots as desired by the operator.

METHODS

To evaluate the effectiveness of this technology, seven surgeons with varying laparoscopic experience tied knots within a simulated minimally invasive setting, using both the automated knot-tying tool and conventional laparoscopic tools. Suture types were 2/0 braided and 4/0 monofilament.

RESULTS

Mean knot-tying times were 246+/-116 s and 102+/-46 s for conventional and automated methods, respectively, showing an average 56% reduction in time per surgeon (p=0.003, paired t-test). The peak holding strength of each knot (the force required to break the suture or loosen the knot) was measured using tensile-testing equipment. These peak holding strengths were normalised by the ultimate tensile strength of each suture type (57.5 N and 22.1N for 2/0 braided and 4/0 monofilament, respectively). Mean normalised holding strengths for all knots were 68.2% and 71.8% of ultimate tensile strength for conventional and automated methods, respectively (p=0.914, paired t-test).

CONCLUSIONS

Experimental data reveal that the automated suturing device has great potential for advancing minimally invasive surgery: it significantly reduced knot-tying times while providing equivalent or greater holding strength than conventionally tied knots.

Для цитирования: Jernigan Shaphan R, Chanoit Guillaume, Veeramani Arun, Owen Stephen B, Hilliard Matthew, Cormier Denis, Laffitte Bryan, Buckner Gregory. A laparoscopic knot-tying device for minimally invasive cardiac surgery.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2009. DOI: 10.1016/j.ejcts.2009.09.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.