DOI 10.2214/ajr.18.20066

Медицинская библиотека · AJR. American journal of roentgenology · 2018

Unexpected Abnormal Uptake in the Breasts at Dedicated Breast PET: Incidentally Detected Small Cancers or Nonmalignant Features?

Satoh Yoko, Motosugi Utaroh, Omiya Yoshie, Onishi Hiroshi

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The purposes of this study were to evaluate the diagnostic performance of dedicated breast PET (dbPET) in cases of unexpected uptake and to define parameters associated with malignancy.

MATERIALS AND

METHODS

There are two types of high-resolution dbPET systems. One has two platelike detectors that compress the breast. This study was conducted with the other type, on which the patient lies prone and the breast hangs through a ring without compression. In total, 627 consecutively registered women underwent 18F-FDG PET/CT and dbPET for pretherapeutic or posttherapeutic evaluation of breast cancer, prior suspicion of breast cancer, or cancer screening. Areas of abnormal FDG uptake, excluding known breast cancers, were analyzed. Uptake was morphologically categorized as focus, mass, or non-mass. Quantitative values were obtained, including the maximum standardized uptake value (SUVmax), peak SUV (SUVpeak), total lesion glycolysis (TLG), metabolic tumor volume (MTV), and lesion-to-background ratio (LBR). Clinical parameters were also assessed. Parameters were compared between breast cancer and noncancer groups; multivariate logistic regression analysis was performed.

RESULTS

Of 40 instances of abnormal uptake, 13 were breast cancer. Morphologic features differed between cancer and noncancer groups (p = 0.0122). Among the cancers, 76.9% (10/13) had mass, 15.4% (2/13) had nonmass, and 7.7% (1/13) had focus uptake. Of noncancerous findings, 3.7% (1/27) were mass, 40.7% (11/27) nonmass, and 55.6% (15/27) focus uptake. SUVpeak (p = 0.0234), TLG (p = 0.0017), MTV (p = 0.004), and LBR (p = 0.0432) also differed between groups. Results of multivariate analysis indicated that morphologic category at dbPET was independently associated with malignancy.

CONCLUSION

Morphologic features of abnormal uptake at dbPET are associated with breast cancer and may be useful for diagnosing lesions of unknown histologic composition.

Для цитирования: Satoh Yoko, Motosugi Utaroh, Omiya Yoshie, Onishi Hiroshi. Unexpected Abnormal Uptake in the Breasts at Dedicated Breast PET: Incidentally Detected Small Cancers or Nonmalignant Features?. AJR. American journal of roentgenology. 2018. DOI: 10.2214/ajr.18.20066.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.