MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/aor.14523

Медицинская библиотека · Artificial organs · 2023

A comprehensive evaluation of hemodynamic energy production and circuit loss using four different ECMO arterial cannulae.

Heinsar Silver, Bartnikowski Nicole, Hartel Gunter, Farah Samia M, Seah E-Peng, Wu Eric, Colombo Sebastiano Maria, Semenzin Clayton, Haymet Andrew, Rätsep Indrek, Pauls Jo, Fraser John F, Suen Jacky Y

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Pulsatile-flow veno-arterial extracorporeal membrane oxygenation (V-A ECMO) has shown encouraging results for microcirculation resuscitation and left ventricle unloading in patients with refractory cardiogenic shock. We aimed to comprehensively assess different V-A ECMO parameters and their contribution to hemodynamic energy production and transfer through the device circuit.

METHODS

We used the i-cor® ECMO circuit, which composed of Deltastream DP3 diagonal pump and i-cor® console (Xenios AG), the Hilite 7000 membrane oxygenator (Xenios AG), venous and arterial tubing and a 1 L soft venous pseudo-patient reservoir. Four different arterial cannulae (Biomedicus 15 and 17 Fr, Maquet 15 and 17 Fr) were used. For each cannula, 192 different pulsatile modes were investigated by adjusting flow rate, systole/diastole ratio, pulsatile amplitudes and frequency, yielding 784 unique conditions. A dSpace data acquisition system was used to collect flow and pressure data.

RESULTS

Increasing flow rates and pulsatile amplitudes were associated with significantly higher hemodynamic energy production (both p < 0.001), while no significant associations were seen while adjusting systole-to-diastole ratio (p = 0.73) or pulsing frequency (p = 0.99). Arterial cannula represents the highest resistance to hemodynamic energy transfer with 32%-59% of total hemodynamic energy generated being lost within, depending on pulsatile flow settings used.

CONCLUSIONS

Herein, we presented the first study to compare hemodynamic energy production with all pulsatile ECLS pump settings and their combinations and widely used yet previously unexamined four different arterial ECMO cannula. Only increased flow rate and amplitude increase hemodynamic energy production as single factors, whilst other factors are relevant when combined.

Для цитирования: Heinsar Silver, Bartnikowski Nicole, Hartel Gunter, Farah Samia M, Seah E-Peng, Wu Eric, Colombo Sebastiano Maria, Semenzin Clayton, Haymet Andrew, Rätsep Indrek, Pauls Jo, Fraser John F, Suen Jacky Y. A comprehensive evaluation of hemodynamic energy production and circuit loss using four different ECMO arterial cannulae.. Artificial organs. 2023. DOI: 10.1111/aor.14523.