MedElevator Войти в исследования
← Поиск публикаций
Медицинская библиотека · The Journal of invasive cardiology · 2014

Effect of caffeine on intravenous adenosine-induced hyperemia in fractional flow reserve measurement.

Matsumoto Hidenari, Nakatsuma Kenji, Shimada Toshihiko, Ushimaru Shunpei, Mikuri Mikiko, Yamazaki Taketoshi, Matsuda Tetsuya

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The interaction between caffeine and adenosine is still a matter of debate.

AIMS: We examined whether caffeine attenuated intravenous adenosine-induced hyperemia in the measurement of fractional flow reserve (FFR) and whether an increased dose of adenosine overcame the caffeine antagonism.

METHODS

FFR was measured using different adenosine doses (140, 175, and 210 μg/kg/min) and papaverine as a reference standard in patients with intermediate coronary stenoses, who refrained from caffeine for >24 h (no-caffeine group; n = 14) and those who consumed caffeine (caffeine group; n = 28).

RESULTS

The median caffeine level in the caffeine group was 2.9 mg/L (interquartile range, 1.8-4.6 mg/L). In the no-caffeine group, FFR with adenosine did not decrease above the dose of 140 μg/kg/min (0.769, 0.771, and 0.770 at 140, 175, and 210 μg/kg/min, respectively) and was not significantly different from that with papaverine (0.765). In the caffeine group, adenosine overestimated FFR (140 μg/kg/min: 0.813, P<.001; 175 μg/kg/min: 0.806, P<.01; 210 μg/kg/min: 0.794, P=.01) compared with papaverine (0.779). The difference in FFR between papaverine and 140 μg/kg/min dose of adenosine was significantly greater in the caffeine group than in the no-caffeine group (0.034 vs 0.004; P<.05).

CONCLUSION

Caffeine attenuates intravenous adenosine-induced hyperemia in FFR measurement. Even increased adenosine doses up to 210 μg/kg/min cannot fully surmount the antagonism.

Для цитирования: Matsumoto Hidenari, Nakatsuma Kenji, Shimada Toshihiko, Ushimaru Shunpei, Mikuri Mikiko, Yamazaki Taketoshi, Matsuda Tetsuya. Effect of caffeine on intravenous adenosine-induced hyperemia in fractional flow reserve measurement.. The Journal of invasive cardiology. 2014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.