DOI 10.1017/s1047951122002864

Медицинская библиотека · Cardiology in the young · 2022

Transcatheter retrieval of atrial septal defect and patent ductus arteriosus occluder: a guidance for device retrieval based on comprehensive bench tests.

Fujii Takanari, Sugiyama Hisashi, Kanazawa Hideaki, Hara Hidehiko, Muneuchi Jun, Yazaki Satoshi

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

The aim of this study is to establish a guidance for device retrieval based on comprehensive bench tests.

BACKGROUND

Device embolisation remains a major complication in transcatheter closure of atrial septal defect and patent ductus arteriosus. Although percutaneous retrieval is feasible in the majority of cases, surgical retrieval may be required in complicated circumstances. However, the methods of transcatheter device retrieval have not been completely established.

METHODS

Bench tests of device retrieval were performed to verify the appropriate retrieval method according to device type/size. The devices used for testing were Amplatzer Septal Occluder (Abbott, Chicago, IL, United States of America), Figulla Flex II (Occlutech GmbH, Jena, Germany), Amplatzer Duct Occluder-I (Abbott), Amplatzer Duct Occluder-II (Abbott), and Amplatzer Vascular Plug-II (Abbott). The retrieval equipment constituted diagnostic catheters (multipurpose catheter and right Judkins catheter, 4-Fr or 5-Fr, Gadelius Medical, Tokyo, Japan), delivery sheath and cables for each device, Amplatz goose neck snares (Medtronic, Minneapolis, MN, United States of America), OSYPKA CATCHER (Osypka ag, Rheinfelden-Herten, Germany), and OSYPKA LASSOS (Osypka). We investigated the retrieval equipment and sheath sizes required for a successful retrieval procedure for variously sized devices.

RESULTS

For patent ductus arteriosus devices, the type of snare and the snaring position are considered important. For atrial septal defect devices, simple snare capture or a double-snare technique with a sufficiently large sheath is effective. Special care should be taken when using the OSYPKA CATCHER for device retrieval.

CONCLUSIONS

The results of this study may assist in the selection of both capture devices and a retrieval sheath or a catheter for complete retrieval.

Для цитирования: Fujii Takanari, Sugiyama Hisashi, Kanazawa Hideaki, Hara Hidehiko, Muneuchi Jun, Yazaki Satoshi. Transcatheter retrieval of atrial septal defect and patent ductus arteriosus occluder: a guidance for device retrieval based on comprehensive bench tests.. Cardiology in the young. 2022. DOI: 10.1017/s1047951122002864.