DOI 10.1177/230949900801600206
Twin hook fixation for proximal femoral fractures.
Аннотация
PURPOSE: To report results of twin hook fixation for proximal femoral fractures in comparison to those fixed with the conventional lag screw.
METHODS
Between August 2005 and July 2006, 2 men and 15 women aged 74 to 94 (mean, 85) years with proximal femoral fractures underwent open reduction and internal fixation using the twin hook system. The tip-apex distance was compared with that in 20 patients treated with the sliding hip screw between August 2004 and July 2005.
RESULTS
In the 17 patients, the hook was inserted into the centre of the femoral head. Bone union was achieved and no intra- or post-operative cut-out or device failure was encountered. In patients using the twin hook and sliding hip screw respectively, the mean tip-apex distance was 22.3 mm and 14.6 mm (p<0.001).
CONCLUSION
Using the twin hook system requires more surgical skill than using the sliding hip screw, because failure to insert the pin into the centre of the femoral head risks intra-articular perforation by the hooks.