Transvenous correction of patent ductus arteriosus in two toy-breed dogs with the Amplatzer™ Vascular Plug 4.
Аннотация
Two Pomeranian dogs referred for interventional correction of a left-to-right shunting patent ductus arteriosus (PDA) had inadequate femoral arterial access for any occlusion device other than micro coils. The decision was made to attempt correction of the PDA using the Amplatzer™ Vascular Plug 4 (AVP4) from a femoral venous approach. An AVP4 was successfully deployed in each dog with complete occlusion noted within 5 min. Complete occlusion was persistent at 24 h after the procedure, while both dogs were subclinical, had no residual ductal flow, and complete or near complete reverse cardiac remodeling at subsequent visits. This report demonstrates the feasibility of PDA occlusion with the AVP4 from the femoral venous approach in small dogs where femoral arterial access is inadequate for other occlusion devices.