Fontan pump: Advanced preclinical development.
Аннотация
OBJECTIVE
Late Fontan failure and attrition is an increasingly prevalent problem for which there is no primary therapy. A Fontan circulation is a circulation that lacks a subpulmonary pump. We are developing a technology that can safely and reliably restore subpulmonary pump function in a Fontan circulation to reverse the Fontan paradox and maintain normal circulatory health.
METHODS
A rotary pump scaled for adolescent and adult use was developed for implantation to physiologically augment flow in the total cavopulmonary connection. Design objectives included 30/70% superior vena cava/inferior vena cava inflow draw over a broad range of performance, and no obstruction in the event of device failure. Hydrodynamic and hemodynamic performance were characterized in vitro in a Fontan mock circulatory loop using blood analog. Hemolysis was characterized in blood.
RESULTS
The pump augments total cavopulmonary connection flow (superior vena cava/inferior vena cava inflow; left pulmonary artery/right pulmonary artery outflow) at +6 to 10 mm Hg, ranging to +18 mm Hg at 3K RPM. Performance is consistent over a wide range of cardiac output. Power requirement is low (1 W) at +6 mm Hg pressure rise. Hemolysis is comparable with other commercially available pumps. Cavitation or suction was not observed. In the stalled condition (0 RPM), gradient is ≤1 mm Hg.
CONCLUSIONS
A Fontan pump can safely reverse the Fontan paradox. Features include a fail-safe contingency that is clinically manageable. The technology may prevent progression of Fontan failure in the setting of preserved systemic ventricular systolic function. As a surgical strategy of long-term biventricular health maintenance, it may improve quality and duration of life for patients with single ventricle heart disease.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.