Analytical performance of the endogenous thrombin potential-based activated protein C resistance assay on the automated ST Genesia system.

Morimont Laure, Leclercq Charline, Didembourg Marie, De Gottal Émilie, Carlo Audrey, Gaspard Ulysse, Dogné Jean-Michel, Douxfils Jonathan

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The evaluation of activated protein C (APC) resistance based on the endogenous thrombin potential (ETP) is recommended during the development of steroid contraceptives in women. In 2019, this assay was validated on the calibrated automated thrombogram (CAT) device. However, in view of its screening potential, its automation is essential.

OBJECTIVES

To transfer the ETP-based APC resistance assay on the ST Genesia system using reagent STG-ThromboScreen with exogenous APC added.

METHOD: Dose-response curves were performed to define APC concentration leading to 90% ETP inhibition on healthy donors. Intra- and interrun reproducibility was assessed. The normal range was defined on the basis of 56 samples from healthy individuals. The sensitivity was assessed on 40 samples from women using combined oral contraceptives (COCs). A method comparison with the validated ETP-based APC resistance on the CAT system was performed. Results were expressed in normalized APC sensitivity ratio (nAPCsr).

RESULTS

The APC concentration leading to 90% ETP inhibition was 652 mU/mL. Intra- and interrun reproducibility showed standard deviation <4%. The nAPCsr normal range stood between 0.00 and 2.20. Analyses of 40 samples from women using COCs confirmed the good sensitivity of the assay. Compared to the CAT system, nAPCsr values were slightly higher on the automated system.

CONCLUSION

This study is the first reporting the analytical performances of the ETP-based APC resistance assay on an automated platform. Results support the concept that this test, when incorporated into clinical routine, could become a promising regulatory and clinical tool to document on the thrombogenicity of female hormonal therapies.

Для цитирования: Morimont Laure, Leclercq Charline, Didembourg Marie, De Gottal Émilie, Carlo Audrey, Gaspard Ulysse, Dogné Jean-Michel, Douxfils Jonathan. Analytical performance of the endogenous thrombin potential-based activated protein C resistance assay on the automated ST Genesia system.. Research and practice in thrombosis and haemostasis. 2022. DOI: 10.1002/rth2.12684.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.