Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL.

Kuwabara Mitsuo, Harada Kanako, Hishiki Yukiko, Kario Kazuomi

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

There is growing evidence of the clinical significance of daytime masked hypertension (MHT) and blood pressure (BP) variability (BPV). Recently, watch-type wearable devices for self-BP measurement have become available. Such devices might be promising tools to identify patients with daytime MHT or large BPV in their real-life conditions. The present study aimed to validate the accuracy of the Omron HEM-6410T-ZM and the Omron HEM-6410T-ZL, which are automatic watch-type wearable devices for self-BP measurement, according to the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 guideline. Watches were held with the wrist at heart level. The mean differences between reference BPs and HEM-6410T-ZM readings were -0.9 ± 7.6/-1.1 ± 6.1 mm Hg for systolic BP (SBP)/diastolic BP (DBP) for criterion 1, and -0.9 ± 6.8/-1.1 ± 5.5 mm Hg for SBP/DBP for criterion 2. The mean differences between reference BPs and HEM-6410T-ZL readings were 2.4 ± 7.3/0.7 ± 7.0 mm Hg for SBP/DBP for criterion 1, and 2.4 ± 6.5/0.7 ± 6.5 mm Hg for SBP/DBP for criterion 2. The Omron HEM-6410T-ZM and the Omron HEM-6410T-ZL both fulfilled both validation criteria 1 and 2 of the ANSI/AAMI/ISO 81060-2:2013 guidelines.

Для цитирования: Kuwabara Mitsuo, Harada Kanako, Hishiki Yukiko, Kario Kazuomi. Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL.. Journal of clinical hypertension (Greenwich, Conn.). 2019. DOI: 10.1111/jch.13499.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.