OBJECTIVE
To determine diazepam, nitrazepam, oxazepam, estazolam, and alprazolam simultaneously in human plasma by reversed phase high-performance liquid chromatography (RP-HPLC).
METHODS
Ten microliter carbamazepine (50 mg/L)as the internal standard was added into 1 mL sample, which contained the 5 mixed sedative hypnotics as standard substance and human plasma as ground substance. They were extracted with acetoacetate from plasma samples, and then were dissolved by 100 microL mobile phase. The blood drug levels were analyzed by high performance liquid chromatograph with 20 microL sample injection on a chromatographic column C18 (4.6 mm x 250 mm) at 30 degree. The mobile phase consisted of methanol and water (65:35),and the flow rate was 1.0 mL/min.The ultraviolet detection wavelength was 230 nm.
RESULTS
The linearity range of the 5 drugs was 5-1,200 microg/L (r> or =0.9966, P<0.05). The recovery rate was 95.5%-105.6%. The extraction recovery rate was more than 75%. The relative standard deviation (RSD) of intra-day and inter-day was less than 10% (n=5).
CONCLUSION
RP-HPLC method is convenient, accurate and sensitive for simultaneous determination of the concentration of diazepam, nitrazepam, oxazepam, estazolam, and alprazolam in human plasma.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.