Медицинская библиотека · The International journal of artificial organs · 2022

Left ventricular venting in veno-arterial extracorporeal membrane oxygenation: A computer simulation study.

Jelenc Matija, Jelenc Blaž, Novak Robert, Poglajen Gregor

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Veno-arterial extracorporeal membrane oxygenation (V-A ECMO) is the fastest way to restore circulation in refractory cardiogenic shock, however it cannot unload the failing left ventricle. There is a lack of consensus regarding optimal approach to left ventricular venting in V-A ECMO patients with severely depressed or absent left ventricular function.

METHODS

A computer model was developed in Matlab Simulink R20016b (MathWorks, Inc., Natick, MA, USA) to analyze different venting options as well as atrial septostomy in the setting of cardiogenic shock and V-A ECMO.

RESULTS

The model has shown an inverse linear relationship between left atrial pressure and either vent, Impella or atrial septum defect flow rate. The minimum vent flow required to prevent pulmonary edema in complete loss of left ventricular function needed to be higher than the bronchial blood flow. Atrial septostomy restored normal pulmonary blood flow with low left atrial pressure but induced stasis in the left ventricle. Venting the pulmonary artery induced stasis in the entire pulmonary circulation as well as left atrium and left ventricle. Venting the left ventricle directly with a cannula or Impella device avoided blood stasis.

CONCLUSION

Our data suggest that reduction of left atrial pressure is linearly related to the vent, Impella or atrial septal defect flow rate. The preferred vent location is the left ventricle as it avoids stasis in the pulmonary circulation and cardiac chambers.

Для цитирования: Jelenc Matija, Jelenc Blaž, Novak Robert, Poglajen Gregor. Left ventricular venting in veno-arterial extracorporeal membrane oxygenation: A computer simulation study.. The International journal of artificial organs. 2022. DOI: 10.1177/03913988221115444.