Diagnostic performance of the Hightop Rapid Diagnostic Test for the detection of anti- Treponema pallidum antibodies in Cameroon: a laboratory-based cross-sectional and comparative study.
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Taheu Ngounouh Christian, Nguwoh Philippe Salomon, Fokam Joseph, Essomba René Ghislain, Halilou Ibrahima, Al-Mayé Bit Younouss Amina, Ibnou Moussa Nafissatou, Yuosembom Emile, Pepa Solange, Mbarga Foe Marcel, Orock Eta Emmanuel, Mpiwouo Panyere Guianni, Madie Tamo Corine, Ngantchui Tchuisseu Jérôme, Moussa Aboubakar, Zotie Elie, Ndombol Tembten John Francois, Nyatte Lydie, Fondoh Victor, Nyanda Nyeng Constance, Taoufick Youssoufa, Nyako Balkissou Dodo, Akenji Mboringong Blaise, Hadja Hamsatou, Okomo Assoumou Marie Claire
Аннотация
DESIGN: A laboratory-based cross-sectional and comparative study was conducted to assess the diagnostic performance of the Hightop Syphilis RDT.
SETTING: Blood samples obtained from adult participants in eight health facilities were analysed at the National Public Health Laboratory (NPHL), Ministry of Public Health, Yaounde, Cameroon.
PARTICIPANTS: From 29 April to 25 August 2023, 583 adult participants of both sexes (aged ≥21 years), including both syphilis positive and syphilis negative, were recruited consecutively in eight health facilities in eight regions of Cameroon.
OUTCOME MEASURES: Blood samples were screened for the detection of anti-Treponema pallidum antibodies using the One Step Rapid Test (Qingdao Hightop Biotech), a non-treponemal test and ELISA (Biorex Diagnostics, UK), a treponemal test used as a reference method. Diagnostic performance of the Syphilis RDT was analysed using Epi Info V.7 and validated through online statistical tools such as StatPages, GraphPad, QuickCalcs and MedCalc software.
RESULTS: Of the 583 samples tested, the Hightop Syphilis RDT revealed a sensitivity of 84.6% (95% CI: 74.8% to 91.1%) and specificity of 98.5% (95% CI: 97.5% to 99.1%). The positive predictive value (PPV) and negative predictive value (NPV) were 84.6% (95% CI: 74.8% to 91.1%) and 98.5% (95% CI: 97.5% to 99.1%), respectively. Regarding the stratification of diagnostic performance by clinical stage, the test showed a sensitivity of 100.0% (95% CI: 71.51% to 100.0%) and specificity of 99.06% (95% CI: 94.86% to 99.98%). The PPV and NPV were 91.67% (95% CI: 61.00% to 98.72%) and 100.0% (95% CI: 96.55% to 100.0%), respectively, in symptomatic individuals. Among asymptomatic individuals, sensitivity was 97.56% (95% CI: 87.14% to 99.94%) and specificity was 100.0% (95% CI: 99.14% to 100.0%). The PPV and NPV were 100.0% (95% CI: 91.19% to 100.0%) and 99.77% (95% CI: 98.40% to 99.97%), respectively.
CONCLUSIONS: The Hightop Syphilis RDT demonstrated adequate diagnostic performance, particularly among symptomatic individuals, supporting its utility as a reliable tool for syphilis detection in clinical settings.