Anatomical reasons for failure of dual-filter cerebral embolic protection application in TAVR: A CT-based analysis.

Voss Stephanie, Campanella Caterina, Burri Melchior, Trenkwalder Teresa, Sideris Konstantinos, Erlebach Magdalena, Ruge Hendrik, Krane Markus, Vitanova Keti, Lange Rüdiger

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The dual-filter Sentinel™ Cerebral Protection System (Sentinel-CPS) is increasingly used during transcatheter aortic valve replacement (TAVR). However, complex vascular anatomy may challenge Sentinel-CPS deployment.

AIM OF THE STUDY: We sought to investigate the impact of anatomic features of the aortic arch and the supra-aortic arteries on technical device failure of Sentinel-CPS application.

METHODS

Analysis of the multislice computed tomography pre-TAVR aortograms of all patients undergoing TAVR with Sentinel-CPS between 2016 and 2020 (n = 92) was performed. We investigated the impact of aortic arch anatomy, configuration, and the angles of the supra-aortic arteries, including the determination of vascular tortuosity index on device failure of Sentinel-CPS application.

RESULTS

The Sentinel-CPS was applied successfully in 83 patients (90.2%). Device failure in nine patients (9.8%) was due to the infeasibility to perform correct deployment of both filters (n = 7) and to obtain peripheral radial access (n = 2). Patients with a failure of Sentinel-CPS application had a higher right subclavian tortuosity index (217 [92-324] vs. 150 [42-252], p = .046), a higher brachiocephalic tortuosity index (27 [5-51] vs. 10 [0-102], p = 0.033) and a larger angulation of the brachiocephalic artery (59° [22-80] vs. 39° [7-104], p = .014) compared with patients with successful application. A brachiocephalic angle more than 59° was predictive for device failure. No differences in aortic arch anatomy or common carotid artery tortuosity were detected between the groups.

CONCLUSIONS

Brachiocephalic tortuosity was found to be associated with failure of Sentinel-CPS application. Filter-based usage should be avoided in TAVR patients with a brachiocephalic angle more than 59°.

Для цитирования: Voss Stephanie, Campanella Caterina, Burri Melchior, Trenkwalder Teresa, Sideris Konstantinos, Erlebach Magdalena, Ruge Hendrik, Krane Markus, Vitanova Keti, Lange Rüdiger. Anatomical reasons for failure of dual-filter cerebral embolic protection application in TAVR: A CT-based analysis.. Journal of cardiac surgery. 2021. DOI: 10.1111/jocs.16025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.