Rupture of extra-corporeal circuit tubing during cardiopulmonary bypass.

Sai Krishna Cheemalapati, Naresh Kumar Palli Venkata, Satpathy Soumya Kanta, Ram Mohan Kanteti, Ramesh Babu Vedangi

Клинический случайСтатьяПеревод в очереди

Аннотация

Roller pumps are widely used for cardiopulmonary bypass in developing nations by virtue of proven safety during several years of institutional use and cost effectiveness. However, careful adjustment of roller occlusion is needed because they are known to cause spallation, tubing wear, and the occasional incident of rupture of tubing in the extracorporeal circuit. Rupture of polyvinylchloride tubing in the pump raceway during repair of a ventricular septal defect in a 4-year-old child is discussed. The event was managed by exclusion and replacement of the defective tubing during a short period of arrest. Use of an inappropriate boot pump and failure to detect its inclusion in the bypass circuit was a significant departure from protocol. However, because occlusion settings and duration of perfusion were within acceptable limits, a manufacturing flaw could also have contributed to tubing failure, and the event may or may not have been averted by the use of larger tubing. In conclusion, this incident reiterates the need for adherence to established protocol during assembly of the pump and draws attention to the fact that tubing integrity is not a guarantee and vigilance is warranted to handle its failure.

Для цитирования: Sai Krishna Cheemalapati, Naresh Kumar Palli Venkata, Satpathy Soumya Kanta, Ram Mohan Kanteti, Ramesh Babu Vedangi. Rupture of extra-corporeal circuit tubing during cardiopulmonary bypass.. The journal of extra-corporeal technology. 2008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.