Multimodal assessment of fetal head circumference and biparietal diameter: systematic differences between MRI and US measurements and development of an US-equivalent prediction regression model.

Ye Kai, Wang Kexin, Wu Pengsheng, Wang Xiaoying

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To investigate the differences between fetal magnetic resonance imaging (MRI) and ultrasonography (US) in measuring head circumference (HC) and biparietal diameter (BPD), and to develop and validate a regression model for predicting US-equivalent values based on MRI measurements.

METHODS

Imaging data of 124 mid-to-late trimester fetuses were retrospectively included. The differences in HC and BPD measurements between the two modalities were quantitatively analyzed. A regression model was constructed using MRI measurements, gestational age (GA), and the time interval between examinations as independent variables to predict the US-equivalent values (USpred).

RESULTS

Compared with US, MRI demonstrated significant systematic overestimation for both HC (mean difference: 14.0 mm, p < 0.001) and BPD (median difference: 3.18 mm, p < 0.001). The established USpred model showed high goodness-of-fit (adjusted R 2 > 0.95). When classifying based on US standards, the USpred values exhibited significantly improved classification efficacy compared to direct MRI measurements: overall accuracy (HC: 0.782 vs. 0.629; BPD: 0.863 vs. 0.782) and Matthews correlation coefficient (HC: 0.512 vs. 0.322; BPD: 0.639 vs. 0.528).

CONCLUSION

Fetal MRI measurements of HC and BPD show a systematic overestimation compared to US. The USpred model can effectively correct for this systematic bias in MRI measurements and the influence of examination time intervals, significantly improving agreement with commonly used clinical US standards. It provides a reliable tool for the precise assessment of prenatal MRI.

Для цитирования: Ye Kai, Wang Kexin, Wu Pengsheng, Wang Xiaoying. Multimodal assessment of fetal head circumference and biparietal diameter: systematic differences between MRI and US measurements and development of an US-equivalent prediction regression model.. Frontiers in radiology. 2026. DOI: 10.3389/fradi.2026.1828351.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.