Intra- and inter-observer agreement in the classification of dyssynergic defecation on high-resolution anorectal manometry using water-perfusion system.
СтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Agrawal Dhiraj, Meshram Megha, Darak Harish, Sadalage Abhishek, Gupta Deepak, Abraham Philip, Bhatia Shobna, Shukla Akash
Аннотация
BACKGROUND AND AIMS: Dyssynergic defecation (DD) classified on anorectal manometry is based on variations in anal sphincter pressures and rectal pressure generated during bearing down. There is a paucity of data on intra- and inter-observer agreement in the classification of DD on high-resolution anorectal manometry (HRAM) using a water-perfusion system.
METHODS: This cross-sectional observational study was conducted between June 2014 and May 2016 on 70 adult patients with DD. Observer 1 did the baseline analysis and reanalyzed the readings at least 6 months later, blinded to the initial analysis results (intra-observer variability). Four other blinded observers (5-25 years in anorectal manometry and minimum 5 years of experience in HRAM) then individually analyzed the manometry data (inter-observer variability).
RESULTS: Intra-observer agreement was excellent (kappa 0.862) for classifying DD. Agreement between the six observations (i.e. two from observer 1 and four from independent observers) was also good (kappa 0.632). The highest agreement was for type 4 DD (kappa 0.738), followed by type 1 (0.680), type 2 (0.664), and type 3 (0.537). Cronbach's alpha value was 0.93, suggesting excellent internal consistency.
CONCLUSION: The diagnosis of the types of DD on anorectal manometry using a water-perfusion system has excellent intra-observer and good inter-observer agreement .
METHODS: This cross-sectional observational study was conducted between June 2014 and May 2016 on 70 adult patients with DD. Observer 1 did the baseline analysis and reanalyzed the readings at least 6 months later, blinded to the initial analysis results (intra-observer variability). Four other blinded observers (5-25 years in anorectal manometry and minimum 5 years of experience in HRAM) then individually analyzed the manometry data (inter-observer variability).
RESULTS: Intra-observer agreement was excellent (kappa 0.862) for classifying DD. Agreement between the six observations (i.e. two from observer 1 and four from independent observers) was also good (kappa 0.632). The highest agreement was for type 4 DD (kappa 0.738), followed by type 1 (0.680), type 2 (0.664), and type 3 (0.537). Cronbach's alpha value was 0.93, suggesting excellent internal consistency.
CONCLUSION: The diagnosis of the types of DD on anorectal manometry using a water-perfusion system has excellent intra-observer and good inter-observer agreement .