Characterization of a CdTe single-photon-counting detector for biomedical imaging applications.

Fardin Luca, Giaccaglia Camilla, Busca Paolo, Bravin Alberto

СтатьяПеревод в очереди

Аннотация

PURPOSE: The Eiger 2X CdTe 1 M-W (Dectris ltd, Baden, Switzerland) single photon counting detector was characterized for imaging applications at the biomedical beamline ID17 of the European Synchrotron Radiation Facility.

METHODS

Linearity, Modulation Transfer Function, Noise Power Spectrum and Detective Quantum Efficiency were measured as a function of photon energy and flux in the range 26-80 keV.

RESULTS

The linearity was confirmed in the flux range specified by Dectris and a detection efficiency higher than 60 % was measured for energies up to 80 keV. The spatial resolution was inferred from the Modulation Transfer Function and was found to be compatible with the pixel size of the detector (75 μm), except at energies just above the K-edge of Cd and Te where it reached 150 μm. The study of the Noise Power Spectrum showed a time-dependency in the response of the sensor, which is mitigated at low photon fluxes (<2⨯108 ph mm-2 s-1).

CONCLUSIONS

This work was the first characterization of the Eiger 2X CdTe 1 M-W for imaging applications with monochromatic synchrotron radiation. The spatial resolution and the quantum efficiency are compatible with low-dose imaging applications.

Для цитирования: Fardin Luca, Giaccaglia Camilla, Busca Paolo, Bravin Alberto. Characterization of a CdTe single-photon-counting detector for biomedical imaging applications.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2023. DOI: 10.1016/j.ejmp.2023.102571.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.