Retrieval of Embolized WATCHMAN® Flex Atrial Appendage Occlusion Device.
Аннотация
This case report documents the management of a 66-year old man with atrial fibrillation with recent placement of a WATCHMAN® Flex atrial appendage occlusion device. The patient presented with renal failure, abdominal pain, and difficulty walking 2 months after placement. The WATCHMAN® Flex device was found to have embolized to his abdominal aorta at the level of the renal arteries with associated thrombus. Extensive workup revealed reduced left ventricular cardiac function and decreased renal function, both of which were felt to be potentially reversible with device removal. The patient then underwent retrieval of the device and all associated thrombus via an open retroperitoneal approach. This case demonstrates a potential consequence of implanting devices such as an atrial appendage occlusion device and describes a technique for removal.