DOI 10.1016/s0886-3350(97)80317-3
Intraocular lens power calculation for microphthalmos.
Аннотация
PURPOSE: To evaluate the refractive results and accuracy of intraocular lens (IOL) power calculation formulas in eyes with microphthalmos.
SETTING
Department of Ophthalmology, Showa University Hospital, Tokyo, Japan.
METHODS
The accuracy of IOL power calculated using the SRK, SRK II, S-SRK, SRK/T, Holladay, and Hoffer Q formulas was evaluated in six eyes with axial lengths less than 19.0 mm.
RESULTS
Postoperative measurement of refraction showed a tendency toward hypermetropia compared with the refraction predicted by each formula. The best predicted refraction was calculated using the SRK/T formula. The tendency for hyperopic estimation was related to the axial length, particularly in eyes with a shorter axial length. However, there was no relationship between the refractive power of the cornea and the error in the predicted refraction by the SRK/T formula. Two eyes with an IOL power of 30.0 diopters (D) had severe hypermetropia.
CONCLUSION
Theoretical formulas were more accurate than empirical ones in eyes with microphthalmos. The severe hypermetropia in the two eyes with a 30.0 D IOL indicates that such patients require a higher IOL power.