Improving the efficacy of sacral nerve stimulation for faecal incontinence by alteration of stimulation parameters.
Аннотация
BACKGROUND
Sacral nerve stimulation (SNS) is an effective treatment for faecal incontinence, but only standard stimulation parameters have been used. This study assessed the clinical impact of altering the parameters.
METHODS
Twelve patients with partially improved faecal incontinence following SNS underwent acute testing to select optimal stimulation parameters; rectal compliance was used as a surrogate marker. Parameters tested were: stimulation off; frequency 14 (standard), 31 or 6.9 Hz; and pulse duration 210 (standard), 450 or 90 micros. Patients completed a 2-week bowel diary, St Mark's continence score (SMCS) and Rockwood faecal incontinence quality-of-life (FIQL) score before testing using standard settings, and after testing using optimized settings.
RESULTS
Optimal settings, determined by greatest increase in rectal compliance, were shorter pulse width in five patients and higher frequency in seven. Optimized stimulation resulted in a decrease in mean episodes of incontinence from 2.3 to 1.2 per week (P = 0.031), soiling from 3.3 to 1.7 days per week (P = 0.016), faecal urgency from 31 to 18 per cent of all evacuations (P = 0.055) and SMCS from 12.3 to 8.7 (P = 0.008); the FIQL coping/behaviour score improved (P = 0.008).
CONCLUSION
With a shorter pulse width and higher frequency, clinical efficacy in patients undergoing SNS for faecal incontinence can be improved.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.