Drugs and brain death diagnostics: determination of drugs capable of inducing EEG zero line.

Meinitzer Andreas, Kalcher Kurt, Gartner Gabriele, Halwachs-Baumann Gabriele, März Winfried, Stettin Mariana

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Several drugs may affect the diagnosis of brain death by depressing the electroencephalographic signal. Serum levels of these drugs must be below their respective therapeutic ranges.

METHODS

A high performance liquid chromatography-based fast and simple method was developed for determination of thiopentone, pentobarbitone, phenobarbitone, methohexital and propofol in serum and validated according to international recommendations.

RESULTS

Separation of extracted analytes was performed on a reversed phase column [Agilent Zorbax SB C18, 5 microm, 4.6 x 250 mm; mobile phase 50% 50 mM NaH(2)PO(4) pH 4.6 mixed with 35% (v/v) acetonitrile and 15% (v/v) methanol]. Calibration curves were linear throughout the selected ranges (microg/mL, thiopentone 0.25-50, pentobarbitone 0.25-25, phenobarbitone 2.5-50, methohexital 0.125-2.50, propofol 0.25-5.0). The standard deviations for the regression line, recovery, imprecision and accuracy results were all highly satisfactory. The lower limits of quantification for propofol, thiopentone and pentobarbitone were set at 0.25 microg/mL, for phenobarbitone 2.5 microg/mL, and for methohexital 0.125 microg/mL, which are below the lowest pharmacologically relevant serum concentrations. Intra- and inter-day coefficients of variation were less than 10% throughout as determined with six replicates.

CONCLUSIONS

The method presented is suitable for drug monitoring to help enhance the reliability of the diagnosis of brain death.

Для цитирования: Meinitzer Andreas, Kalcher Kurt, Gartner Gabriele, Halwachs-Baumann Gabriele, März Winfried, Stettin Mariana. Drugs and brain death diagnostics: determination of drugs capable of inducing EEG zero line.. Clinical chemistry and laboratory medicine. 2008. DOI: 10.1515/cclm.2008.335.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.