DOI 10.1016/j.rvsc.2008.02.003
Evaluation the Clinitek status automated dipstick analysis device for semiquantitative testing of canine urine.
Аннотация
The aim of this prospective study was to evaluate the Clinitek status analyser using Multistix10SG/Microalbustix dipsticks (all: Siemens Dx) for canine urine (n=101) compared to reference methods: visual reading (Combur9 dipstick, Roche), refractometry, microscopy, and quantitative protein/creatinine analysis (Pentra400, AxonLab). The automated analyses were done twice and visual tests were performed by two examiners. An excellent to good concordance was demonstrated between the first/second analysis with the Multistix10SG and the Combur9 dipstick, respectively with Cohen's kappa-values ranging from 0.776 to 1.000. Agreement between both dipsticks was good for glucose (kappa=0.753), blood (kappa=0.793), protein (kappa=0.788), and moderate for bilirubin (kappa=0.431) and ketones (kappa=0.540). In 6/101 specimens, false positive ketone reactions were obtained with the Multistix10SG. Multistix10SG could not be used for determination of pyuria or specific gravity. Semiquantitative/quantitative protein results correlated well (rho=0.90) and creatinine measurements moderately (rho=0.76). Due to automated data transmission to the laboratory information system, the Clinitek status is of advantage in veterinary laboratories/clinics.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.