MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.bot.0000177106.30837.28

Медицинская библиотека · Journal of orthopaedic trauma · 2005

Computer-assisted guidewire insertion for hip fracture fixation.

Mayman D, Vasarhelyi E M, Long W, Ellis R E, Rudan J, Pichora D R

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

: This study was designed to test in a laboratory setting a novel computer-assisted fluoroscopic technique and a conventional fluoroscopic technique for open reduction and internal fixation (ORIF) of hip fractures. Our hypothesis is that a novel computer-assisted fluoroscopic technique will achieve acceptable guidewire placement in one pass, with decreased fluoroscopic time and with accuracy and precision better than conventional technique.

DESIGN

Prospective, randomized trials.

SETTING

Laboratory.

PARTICIPANTS

Thirty, Sawbone, femur phantoms.

INTERVENTION: Dynamic hip screw guidewires were inserted into 15 femur phantoms under fluoroscopic guidance by using computer-assisted fluoroscopic ORIF technique, and 15 femurs were inserted by using a conventional fluoroscopic-assisted ORIF technique.

MAIN OUTCOME MEASUREMENTS: Ideal guidewire placement was defined as the center of the femoral head, 5 mm from the apical bone edge on anteroposterior and lateral views. Accuracy was measured as distance to ideal placement, and the number of passes and fluoroscopic time were noted for each trial.

RESULTS

The computer-assisted technique achieved an average guidewire placement that was as accurate as the conventional technique in fewer passes, 1.1 +/- 0.2 (mean +/- standard deviation) compared with 2.4 +/- 1.1 (P < 0.0001), respectively, and with fewer fluoroscopic images, 2 +/- 0 compared with 13.5 +/- 3 (P < 0.0002), respectively. Guidewire placement in both groups was within the tip-apex distance defined by Baumgaertner et al.

CONCLUSIONS

The computer-assisted technique was significantly more accurate and precise than conventional technique. It also required fewer drill tracks through the femur and exposed the patient and the surgical team to significantly less ionizing radiation.

Для цитирования: Mayman D, Vasarhelyi E M, Long W, Ellis R E, Rudan J, Pichora D R. Computer-assisted guidewire insertion for hip fracture fixation.. Journal of orthopaedic trauma. 2005. DOI: 10.1097/01.bot.0000177106.30837.28.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.