Aerosol and droplet generation from mandible and midface fixation: Surgical risk in the pandemic era.

Ye Michael J, Sharma Dhruv, Campiti Vincent J, Rubel Kolin E, Burgin Sarah J, Illing Elisa A, Ting Jonathan Y, Park Jae Hong, Johnson Jeffrey D, Vernon Dominic J, Lee Hui Bae, Nesemeier B Ryan, Shipchandler Taha Z

СтатьяПеревод в очереди

Аннотация

PURPOSE: The COVID-19 pandemic has led to concerns over transmission risk from healthcare procedures, especially when operating in the head and neck such as during surgical repair of facial fractures. This study aims to quantify aerosol and droplet generation from mandibular and midface open fixation and measure mitigation of airborne particles by a smoke evacuating electrocautery hand piece.

MATERIALS AND

METHODS

The soft tissue of the bilateral mandible and midface of two fresh frozen cadaveric specimens was infiltrated using a 0.1% fluorescein solution. Surgical fixation via oral vestibular approach was performed on each of these sites. Droplet splatter on the surgeon's chest, facemask, and up to 198.12 cm (6.5 ft) away from each surgical site was measured against a blue background under ultraviolet-A (UV-A) light. Aerosol generation was measured using an optical particle sizer.

RESULTS

No visible droplet contamination was observed for any trials of mandible or midface fixation. Total aerosolized particle counts from 0.300-10.000 μm were increased compared to baseline following each use of standard electrocautery (n = 4, p < 0.001) but not with use of a suction evacuating electrocautery hand piece (n = 4, p = 0.103). Total particle counts were also increased during use of the powered drill (n = 8, p < 0.001).

CONCLUSIONS

Risk from visible droplets during mandible and midface fixation is low. However, significant increases in aerosolized particles were measured after electrocautery use and during powered drilling. Aerosol dispersion is significantly decreased with the use of a smoke evacuating electrocautery hand piece.

Для цитирования: Ye Michael J, Sharma Dhruv, Campiti Vincent J, Rubel Kolin E, Burgin Sarah J, Illing Elisa A, Ting Jonathan Y, Park Jae Hong, Johnson Jeffrey D, Vernon Dominic J, Lee Hui Bae, Nesemeier B Ryan, Shipchandler Taha Z. Aerosol and droplet generation from mandible and midface fixation: Surgical risk in the pandemic era.. American journal of otolaryngology. 2020. DOI: 10.1016/j.amjoto.2020.102829.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.