Peak velocity and flow quantification validation for sensitivity-encoded phase-contrast MR imaging.

Lew Calvin D, Alley Marcus T, Bammer Roland, Spielman Daniel M, Chan Frandics P

Сравнительное исследованиеСтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

RATIONALE AND

OBJECTIVES

Phase-contrast (PC) magnetic resonance imaging (MRI) technique has important clinical applications in blood flow quantification and pressure gradient estimation by velocity measurement. Parallel imaging using sensitivity encoding (SENSE) may substantially reduce scan time. We demonstrate the utility of PC-MRI measurements accelerated by SENSE under clinical conditions.

MATERIALS AND

METHODS

Accuracy and repeatability of a SENSE-PC implementation was evaluated by comparison with a commercial PC sequence with five normal volunteers. Twenty-six patients were then scanned with SENSE-PC at reduction factors (R = 1, 2, and 3). Blood flow and peak velocity were measured in the aorta and pulmonary trunk in 16 patients and peak velocity was measured at the coarctation of 10 patients. Quantitative flow, shunt ratio, and peak velocity measurements obtained with different reduction factors were compared using correlation, linear regression, and Bland-Altman statistics. All studies were approved by an Institutional Review Board, and informed consent was acquired from all subjects.

RESULTS

The correlation coefficients for all comparisons were >0.962 and with high statistical significance (P < .01). Linear regression slopes ranged between 0.96 and 1.11 for flow and 0.88 to 1.05 for peak velocity. For flow, the Bland-Altman statistics yielded a total mean difference ranging from -0.02 to 0.05) L/minute with 2 standard of deviation limits ranging from -0.52 to 0.75 L/minute. For peak velocity, the total mean difference ranged from -0.10 to -0.004) milliseconds with 2-SD limits ranging from -0.062 to 0.46 milliseconds. R = 3 to R = 1 comparisons had greater 2-SD limits than R = 2 to R = 1 comparisons.

CONCLUSION

SENSE PC-MRI measurements for flow and pressure gradient estimation were comparable to conventional PC-MRI.

Для цитирования: Lew Calvin D, Alley Marcus T, Bammer Roland, Spielman Daniel M, Chan Frandics P. Peak velocity and flow quantification validation for sensitivity-encoded phase-contrast MR imaging.. Academic radiology. 2007. DOI: 10.1016/j.acra.2006.11.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.