DOI 10.1136/archdischild-2020-320531

Медицинская библиотека · Archives of disease in childhood. Fetal and neonatal edition · 2021

Comparison of inspiratory effort with three variable-flow nasal continuous positive airway pressure devices in preterm infants: a cross-over study.

Bordessoule Alice, Moreira Amelia, Felice Civitillo Cristina, Combescure Christophe, Polito Angelo, Rimensberger Peter C

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Patient's work of breathing may vary between different neonatal nasal continuous positive airway pressure (NCPAP) devices. Therefore, we aimed to compare the inspiratory effort of three variable-flow NCPAP delivery systems used in preterm infants.

DESIGN

Cross-over study.

PATIENTS/

SETTING

From June 2015 to August 2016, 20 preterm infants weighing ≤2500 g requiring NCPAP for mild respiratory distress syndrome were enrolled.

INTERVENTIONS: Each patient was successively supported by three randomly assigned variable-flow NCPAP systems (MedinCNO, Infant Flow and Servo-i) for 20 min while maintaining the same continuous positive airway pressure level as the patient was on before the study period.

MAIN OUTCOME MEASURES: Patients' inspiratory effort was estimated by calculating the sum of the difference between maximal inspiratory and baseline electrical activity of the diaphragm (∆EAdi) for 30 consecutive breaths, and after normalising this obtained value for the timing of the 30 breaths.

RESULTS

Physiological parameters (oxygen saturation measured by pulse oximetry, respiratory rate, heart beat, transcutaneous partial pressure CO2) and oxygen requirements remained identical between the three NCPAP systems. Although a wide variability in inspiratory effort could be observed, there were no statistically significant differences between the three systems for the sum of ∆EAdi for 30 breaths: CNO, 262 (±119) µV; IF, 352 (±262) µV; and SERVO-i, 286 (±126) µV, and the ∆EAdi reported on the timing of 30 breaths (sum ∆EAdi/s): CNO, 6.1 (±2.3) µV/s; IF, 7.9 (±4.9) µV/s; SERVO-i, 7.6 (±3.6) µV/s.

CONCLUSION

In a neonatal population of preterm infants, inspiratory effort is comparable between the three tested modern variable-flow NCPAP devices.

Для цитирования: Bordessoule Alice, Moreira Amelia, Felice Civitillo Cristina, Combescure Christophe, Polito Angelo, Rimensberger Peter C. Comparison of inspiratory effort with three variable-flow nasal continuous positive airway pressure devices in preterm infants: a cross-over study.. Archives of disease in childhood. Fetal and neonatal edition. 2021. DOI: 10.1136/archdischild-2020-320531.