m+Medelevator.RU
← Поиск публикаций
The Journal of urology · 2006

Ultrasound assessment of detrusor thickness in men-can it predict bladder outlet obstruction and replace pressure flow study?

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kessler Thomas M, Gerber Rolf, Burkhard Fiona C, Studer Urs E, Danuser Hansjörg

Аннотация

PURPOSE: We estimated the diagnostic accuracy of ultrasound detrusor thickness measurement for BOO and investigated whether this method can replace PFS for the diagnosis of BOO in some patients with lower urinary tract symptoms.

MATERIALS AND METHODS: Detrusor thickness was measured by linear ultrasound (7.5 MHz) at a filling volume of greater than 50% of cystometric capacity in 102 men undergoing PFS for LUTS. All patients with prior treatment for bladder outlet obstruction and those with underlying neurological disorders were excluded from analysis. Detrusor thickness was correlated with PFS data. Obstruction was defined according to the Abrams-Griffiths nomogram.

RESULTS: Detrusor thickness was significantly higher (p <0.0001) in obstructed (61 cases, median detrusor thickness 2.7 mm, IQR 2.4 to 3.3) compared to unobstructed (18 cases, median detrusor thickness 1.7 mm, IQR 1.5 to 2) as well as equivocal (23 cases, median detrusor thickness 1.8 mm, IQR 1.5 to 2.2) cases. A weak to medium Spearman correlation was found between detrusor thickness and PFS parameters. For a diagnosis of BOO, detrusor thickness of 2.9 mm or greater had a positive predictive value of 100%, a negative predictive value of 54%, specificity of 100% and sensitivity of 43%. ROC analysis revealed that detrusor thickness had a high predictive value for BOO with an AUC of 0.88 (95% CI 0.81-0.94).

CONCLUSIONS: In men with LUTS without prior treatment and/or neurological disorders, ultrasonographically assessed detrusor thickness 2.9 mm or greater has a high predictive value for BOO and can replace PFS for the diagnosis of BOO. However, this cutoff value needs to be validated in a larger study population.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты