Efficacy of Wearable Single-Lead ECG Monitoring during Exercise Stress Testing: A Comparative Study.
Аннотация
BACKGROUND AND
OBJECTIVES
Few comparative studies have evaluated wearable single-lead electrocardiogram (ECG) devices and standard multi-lead ECG devices during exercise testing. This study aimed to validate the accuracy of a wearable single-lead ECG monitor for recording heart rate (HR) metrics during graded exercise tests (GXTs).
METHODS
A cohort of 50 patients at a tertiary hospital underwent GXT while simultaneously being equipped with wearable single- and conventional multi-lead ECGs. The concordance between these modalities was quantified using the intraclass correlation coefficient and Bland-Altman plot analysis.
RESULTS
The minimum and average HR readings between the devices were generally consistent. Parameters such as ventricular ectopic beats and supraventricular ectopic beats showed strong agreement. However, the agreement for the Total QRS and Maximum RR was not sufficient. HR measurements across different stages of the exercise test showed sufficient agreement. Although not statistically significant, the standard multi-lead ECG devices exhibited higher noise levels compared to the wearable single-lead ECG devices.
CONCLUSIONS
Wearable single-lead ECG devices can reliably monitor HR and detect abnormal beats across a spectrum of exercise intensities, offering a viable alternative to traditional multi-lead systems.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.