MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11548-015-1296-9

Медицинская библиотека · International journal of computer assisted radiology and surgery · 2015

Vision-based real-time position control of a semi-automated system for robot-assisted joint fracture surgery.

Dagnino Giulio, Georgilas Ioannis, Tarassoli Payam, Atkins Roger, Dogramadzi Sanja

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: Joint fracture surgery quality can be improved by robotic system with high-accuracy and high-repeatability fracture fragment manipulation. A new real-time vision-based system for fragment manipulation during robot-assisted fracture surgery was developed and tested.

METHODS

The control strategy was accomplished by merging fast open-loop control with vision-based control. This two-phase process is designed to eliminate the open-loop positioning errors by closing the control loop using visual feedback provided by an optical tracking system. Evaluation of the control system accuracy was performed using robot positioning trials, and fracture reduction accuracy was tested in trials on ex vivo porcine model.

RESULTS

The system resulted in high fracture reduction reliability with a reduction accuracy of 0.09 mm (translations) and of [Formula: see text] (rotations), maximum observed errors in the order of 0.12 mm (translations) and of [Formula: see text] (rotations), and a reduction repeatability of 0.02 mm and [Formula: see text].

CONCLUSIONS

The proposed vision-based system was shown to be effective and suitable for real joint fracture surgical procedures, contributing a potential improvement of their quality.

Для цитирования: Dagnino Giulio, Georgilas Ioannis, Tarassoli Payam, Atkins Roger, Dogramadzi Sanja. Vision-based real-time position control of a semi-automated system for robot-assisted joint fracture surgery.. International journal of computer assisted radiology and surgery. 2015. DOI: 10.1007/s11548-015-1296-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.