MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/bja/53.4.351

Медицинская библиотека · British journal of anaesthesia · 1981

Non-invasive measurement of cardiac output during anaesthesia. An evaluation of the soluble gas uptake method.

Heneghan C P, Branthwaite M A

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

A technique for measuring cardiac output which depends on the uptake of an inert soluble gas from the lungs has been evaluated during anesthesia. A respiratory mass spectrometer has been used to follow the concentrations of argon and freon-22 during passive rebreathing in anaesthetized patients before cardiopulmonary bypass. Values for cardiac output obtained with this technique are reproducible, but lower than those recorded using the direct Fick technique before and after the rebreathing manoeuvre. A reduction in cardiac output caused by vigorous rebreathing is the most likely explanation for the discrepancy and, although serial measurements of oxygen consumption may permit application of a correction factor, a method of measurement which causes significant haemodynamic disturbance cannot be recommended for widespread use.

Для цитирования: Heneghan C P, Branthwaite M A. Non-invasive measurement of cardiac output during anaesthesia. An evaluation of the soluble gas uptake method.. British journal of anaesthesia. 1981. DOI: 10.1093/bja/53.4.351.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.