The Gore Cardioform Atrial Septal Defect Occluder: A novel solution to the management of severe hemolysis following transcatheter septal defect closure.

Naimi Iman, Jones Thomas K, Steinberg Zachary

Клинический случайСтатьяПеревод в очереди

Аннотация

Significant hemolysis is a recognized complication of transcatheter high-velocity shunt occlusion using some Amplatzer devices. We describe a case of severe hemolysis following occlusion of an iatrogenic Gerbode defect with an Amplatzer muscular ventricular septal defect occluder successfully managed by transcatheter device removal and reocclusion with a Gore Cardioform Atrial Septal Defect Occluder.

Для цитирования: Naimi Iman, Jones Thomas K, Steinberg Zachary. The Gore Cardioform Atrial Septal Defect Occluder: A novel solution to the management of severe hemolysis following transcatheter septal defect closure.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2022. DOI: 10.1002/ccd.30144.