Intermittently Scanned Continuous Glucose Monitoring Performance in Patients With Liver Cirrhosis.

Rigon Fernanda Augustini, Ronsoni Marcelo Fernando, Hohl Alexandre, Vianna André Gustavo Daher, Sande-Lee Simone van de, Schiavon Leonardo de Lucca

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

AIM: To evaluate the use of intermittently scanned continuous glucose monitoring (isCGM) in patients with liver cirrhosis (LC).

METHODS

Observational study including 30 outpatients with LC (Child-Pugh B/C): 10 without diabetes (DM) (G1), 10 with newly diagnosed DM by oral glucose tolerance test (G2), and 10 with a previous DM diagnosis (G3). isCGM (FreeStyle Libre Pro) was used for 56 days (four sensors/patient). Blood tests were performed at baseline and after 28 and 56 days.

RESULTS

No differences were found in the baseline characteristics, except for higher age in G3. There were significant differences between G1, G2 and G3 in glucose management indicator (GMI) (5.28 ± 0.17, 6.03 ± 0.59, 6.86 ± 1.08%, P < .001), HbA1c (4.82 ± 0.39, 5.34 ± 1.26, 6.97 ± 1.47%, P < .001), average glucose (82.79 ± 7.06, 113.39 ± 24.32, 149.14 ± 45.31mg/dL, P < .001), time in range (TIR) (70.89 ± 9.76, 80.2 ± 13.55, 57.96 ± 17.96%, P = .006), and glucose variability (26.1 ± 5.0, 28.21 ± 5.39, 35.31 ± 6.85%, P = .004). There was discordance between GMI and HbA1c when all groups were considered together, with a mean difference of 0.35% (95% SD 0.17, 0.63). In G1, the mean difference was 0.46% (95% SD 0.19, 0.73) and in G2 0.69% (95% SD 0.45, 1.33). GMI and HbA1c were concordant in G3, with a mean difference of -0.10 % (95% SD [-0.59, 0.38]).

CONCLUSION

Disagreements were found between the GMI and HbA1c levels in patients with LC. isCGM was able to detect abnormalities in glycemic control that would not be detected by monitoring with HbA1c, suggesting that isCGM can be useful in assessing glycemic control in patients with LC.

Для цитирования: Rigon Fernanda Augustini, Ronsoni Marcelo Fernando, Hohl Alexandre, Vianna André Gustavo Daher, Sande-Lee Simone van de, Schiavon Leonardo de Lucca. Intermittently Scanned Continuous Glucose Monitoring Performance in Patients With Liver Cirrhosis.. Journal of diabetes science and technology. 2024. DOI: 10.1177/19322968241232686.