Five-year Outcomes With Conformable GORE TAG Endoprosthesis Used in Traumatic Aortic Transections.

Farber Mark A, Krishnasastry Kambhampaty V, Desai Nimesh, Starnes Benjamin W, Matsumura Jon S, Tohill Beth C, TAG 08-02 clinical trial investigators

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The Conformable GORE TAG thoracic endoprosthesis (CTAG) device (W.L. Gore and Associates, Flagstaff, AZ) seeks to optimize thoracic endovascular repair of blunt aortic injuries by better apposition to the aortic arch. The Evaluation of the GORE Conformable TAG Thoracic Endoprosthesis for Treatment of Traumatic Transection (TAG 08-02) study assessed the safety and effectiveness of the CTAG device in the 5 years after treatment.

METHODS

This prospective, single-arm study was conducted at 30 US sites with follow-up through 5 years. Patients were evaluated for all-cause mortality and device events through follow-up. An independent core laboratory reviewed pretreatment and posttreatment images.

RESULTS

A total of 51 initial cohort patients and 50 continued access patients were enrolled and underwent placement of a total of 111 CTAG devices (mean, 1.1/subject; range, 1.0 to 3.0) during initial treatment. On average, patients were 42.7 ± 19.5 years old, with a mean Injury Severity Score of 31.5 ± 14.5, and most presented with polytraumatic injuries. The median time between injury and treatment was 21.0 hours (range, 3.2 to 334.4 hours). A total of 60 patients had partial or complete left subclavian artery coverage, but only 1 patient had a stroke that was unable to be attributed to the device or procedure. The freedom from all-cause mortality was 95% and 89% at 1 month and 5 years after the procedure, respectively. There were 2 minor endoleaks. No aortic ruptures, wire frame fractures, erosions, lumen obstructions, device compressions, or thrombus-related events were reported.

CONCLUSIONS

The 5-year outcomes verify that the CTAG device is a safe, effective, and durable option for patients with blunt aortic injuries who are undergoing thoracic endovascular aortic repair.

Для цитирования: Farber Mark A, Krishnasastry Kambhampaty V, Desai Nimesh, Starnes Benjamin W, Matsumura Jon S, Tohill Beth C, TAG 08-02 clinical trial investigators. Five-year Outcomes With Conformable GORE TAG Endoprosthesis Used in Traumatic Aortic Transections.. The Annals of thoracic surgery. 2021. DOI: 10.1016/j.athoracsur.2021.05.054.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.